Re: Afnor Viscosity cups


[ Follow Ups/Ãß°¡ÀÇ°ß ¿­¶÷ ] [ Post Followup/Ãß°¡ÀÇ°ß Á¦½Ã ] [ Technical Consultation - Flow Measurement Technology ]

Posted by(ÀÛ¼ºÇϽŠºÐ) MaxMet Incorporation ( December 07, 2000 at 12:13:11: )

In Reply to(¿ø¹®) : Afnor Viscosity cups
posted by(¿ø¹® ÀÛ¼ºÀÚ) : L; Rucquoij on December 04, 2000 at 22:59:03:

: Hi,
: Can annyone gife me specifications ( or a formula for calulating)on AFNOR VISCOSOTY CUPS regardig run-true times vs viscosity ?
: Thank you

=======================================
This is MaxMet's internal data.

MaxMet takes no responsibility for your
non-commercial or commercial use of the data.


================================================
COUPE CONSISTOMETRIQUE 2,5 NF T30-014
(for 2.5 mm orifice AFNOR cup)

Empirical Correlation * : KV = 0.0002 T2 + 0.6243 T - 12.77
KV : kinematic viscosity (mm2/s)
T : AFNOR second (s)uncertainty : less than ¡¾ 2%
DATA by MaxMet Research Institute, copyright MaxMet Inc. 2000
==============================================================

MaxMet waives the copyright reservations
of its data for non-commercial use
of the data (e.g., FREE publication by
Universities, Governments, NGOs etc.).
However you may not sell or use it as
commerical purpose without our approval.